Test Information & Changes

Open
Lead Testing Update! Additional Important Information!
5/19/2017 11:35:25 AM

Dear Healthcare Provider,

Centers for Disease Control has issued an Advisory related to lead blood testing. It is now believed that the vacutainer used is not at fault as previously communicated by the FDA. It is now being communicated that the LeadCare analyzer should no longer be used with venous blood samples.

In light of this new information, we have changed our procedure for lead testing and will be sending samples to ARUP.  Please read the details here regarding specimen collection and the acceptable vacutainers for lead testing sample collection and review the CDC Advisory for additional information.

Thank you,
PCL Alverno

 

URGENT UPDATE RELATED TO BLOOD LEAD COLLECTION!
5/16/2017 4:42:25 PM

Dear Healthcare Provider,

The FDA has issued a strong recommendation to immediately discontinue the use of BD Lavender and Tan Top EDTA vacuntainer tubes for blood lead testing.  Please read this for more details on the issue and which tubes will be acceptable for lead testing.

Thank you!

PCL Alverno May 2017 Test Bulletin
5/3/2017 12:56:13 PM

Dear Healthcare Provider,

The May 2017 Test Bulletin is available for your review. It includes news on page one regarding CA 15-3 testing. If you order this test in your practice, please check page one of the test bulletin.

Page two includes information about the replacement of antiphospholipid antibody testing with the new antiphospholipid syndrome reflex panel, effective June 1, 2017. 

For more information on the new antiphospholipid syndrome reflex panel, visit ARUP Connect 

Page 3 provides you information on ROS1 by FISH testing related to non-small cell lung cancer.

We honor our fallen soldiers on Memorial Day, May 29, 2017